Views: 0 Author: Site Editor Publish Time: 2026-11-04 Origin: Site
Pressure transmitters are widely used in chemical plants, power stations, water treatment systems and other industrial processes. They measure process pressure and send signals to a PLC or DCS for monitoring and control.
Like other industrial instruments, pressure transmitters may develop faults during operation. Common problems include no display, no output, signal drift, unstable readings, inaccurate measurement and communication failure.
Many of these problems are related to power supply, wiring, process connections, installation or calibration.
This guide explains six common pressure transmitter faults, their possible causes and practical troubleshooting steps.
The LCD display is blank.
The 4–20mA output is zero or abnormally low.
The control system shows a fault or invalid value.
Incorrect power supply voltage
Reversed polarity or loose wiring
Damaged signal cable
Excessive loop load resistance
Faulty display module or electronic circuit
Check the power supply. Measure the voltage directly at the transmitter terminals. Make sure it meets the manufacturer's requirements.
Inspect the wiring. Check polarity, terminal connections and cable condition.
Check the loop resistance. Make sure the total loop load is within the permitted operating range.
Check the display. If the display is blank but the output current is normal, the display module may be faulty.
Inspect the electronics. If the external connections and power supply are normal, the transmitter may need professional inspection.
Tip: If an isolator or safety barrier is installed, check the voltage at the transmitter terminals rather than relying only on the power supply reading in the control cabinet.
The process pressure remains stable, but the output slowly changes.
The output does not return to 4mA after the pressure is released to zero.
Output drift can be divided into two types:
Zero drift: The output is offset from its expected zero-pressure value.
Span drift: The output changes at an incorrect rate across the measuring range.
Blocked or partially blocked impulse lines
Liquid or deposits inside the pressure line
Diaphragm contamination or damage
Temperature changes
Mechanical stress from installation
Long-term component aging
Check the impulse line for blockage, leakage or liquid accumulation.
Inspect the process diaphragm for deposits or damage.
Check whether temperature changes affect the measurement.
Release the process pressure safely and check the zero output.
If necessary, perform zero and span calibration according to the manufacturer's procedure.
If the output drifts again shortly after calibration, investigate the process connection, impulse line and diaphragm before repeatedly adjusting the transmitter.
The output current changes rapidly.
The DCS trend shows irregular fluctuations.
The process pressure appears stable, but the transmitter reading keeps jumping.
First, determine whether the pressure is actually fluctuating or whether the instrument signal is unstable.
If several instruments on the same process line show similar fluctuations, the process pressure itself may be changing. If only one transmitter is affected, check the instrument and its wiring.
Electromagnetic interference from motors or variable frequency drives
Poor cable shielding or grounding
Loose or corroded terminals
Mechanical vibration
Actual process pressure pulsation
Inspect the grounding and cable shielding according to the instrument manual and site wiring requirements.
Check terminal connections for looseness or corrosion.
Keep signal cables away from power cables where possible.
Check the mounting bracket and surrounding vibration sources.
Compare the reading with another pressure instrument on the same process line.
Adjust the damping setting only when appropriate for the process response.
Increasing damping may reduce visible fluctuations, but it does not solve the underlying cause.
The output is stable but does not match a reference pressure instrument.
The measurement remains consistently higher or lower than the actual process pressure.
The error increases across the measuring range.
Blocked, leaking or incorrectly installed impulse lines
Incorrect installation or process connection
Damaged diaphragm
Incorrect zero or range settings
Calibration error
Instrument performance outside its specified tolerance
Compare the readings. Use a suitable calibrated pressure reference under the same pressure conditions.
Inspect the impulse line. Check for blockage, leakage, trapped liquid and valve problems.
Check the installation. Inspect the process connection, sealing and mounting condition.
Verify the zero point. Check the output under the correct zero-pressure condition.
Check the measuring range. Confirm that the configured lower and upper range values are correct.
Perform calibration. Test multiple pressure points across the measuring range.
If the transmitter remains outside its specified tolerance after proper checks and adjustment, arrange professional verification or repair.
The process pressure changes, but the output current remains almost constant.
The DCS reading stays at one value.
A closed pressure isolation valve
A blocked impulse line
A damaged or stuck diaphragm
Incorrect range configuration
Electronic circuit failure
Check whether the isolation valve is fully open.
Inspect the impulse line for blockage.
Confirm that the actual process pressure is changing by comparing it with a suitable local pressure gauge or reference instrument.
Check whether the transmitter range is configured correctly.
If the pressure reaches the sensing element but the output does not respond, arrange further inspection.
Do not assume the transmitter is faulty before confirming that process pressure can reach its sensing element.
The output reaches or exceeds its configured upper or lower alarm limit.
The transmitter reports a sensor or diagnostic alarm.
A HART communicator cannot connect or communication is intermittent.
Process pressure exceeds the configured measuring range.
The sensor or diaphragm is damaged.
The configured range does not match the actual process conditions.
The loop power supply or wiring has a fault.
Insufficient loop resistance
Insufficient voltage at the transmitter terminals
Loose wiring or poor electrical connections
Electrical interference
Incompatible communicator or device configuration
Measure the loop voltage and output current.
Check the transmitter's configured measuring range, units and output mode.
Confirm that the actual process pressure is within the rated range.
For HART communication, check the loop resistance and terminal voltage against the transmitter and communicator manuals.
Inspect the wiring and grounding for loose connections or interference.
If the problem remains, check device compatibility and arrange technical inspection.
Important: A 250Ω loop resistance is commonly required for HART communication, but the permitted total loop resistance depends on the power supply and transmitter design. Always check the applicable manual.
Regular inspection can reduce unexpected failures and improve measurement reliability.
Check impulse lines regularly. Look for blockage, leakage and liquid accumulation.
Inspect wiring and grounding. Keep signal cables properly installed and protected from electrical interference.
Check the installation. Make sure the process connection and mounting are suitable for the application.
Follow a calibration schedule. Verify the instrument according to the applicable procedures and site requirements.
Select the correct measuring range. Avoid choosing a range that is unsuitable for normal operating pressure or possible pressure peaks.
Common causes include insufficient power, incorrect wiring, excessive loop load or an internal electronic fault.
Check for process pressure pulsation, electrical interference, poor grounding, loose wiring and mechanical vibration.
Possible causes include zero drift, trapped pressure, a blocked impulse line or incorrect zero configuration. Check the actual zero-pressure condition before adjustment.
The calibration interval depends on the application, operating conditions, instrument accuracy requirements and maintenance plan.
Yes. Many transmitters can provide a 4–20mA output without a local display. The display is an optional indication feature on some models.
Most pressure transmitter faults can be traced to four main areas: power supply, process connection, signal transmission and calibration.
A systematic troubleshooting process helps identify the actual cause before replacing or adjusting the instrument. Always follow the manufacturer's instructions and site safety procedures when inspecting or servicing a pressure transmitter.
Looking for a reliable pressure measurement solution? Explore JIECHUANG Pressure Transmitters or contact our team for product selection and technical support.